Chat or Call
吳燕紅
Sales Manager - 吳燕紅
姚益涵
Sales Manager - 姚益涵
吳燕紅
Sales Manager - 吳燕紅
姚益涵
Sales Manager - 姚益涵
姚益涵
Sales Manager - 姚益涵
Select language
4 Ranges Coin Cell Battery testing System
CT-4008Q-5V100mA
  • Size of Each Unit (W*D*H)(mm)
    1U(19"),483.6*321.5*48.5 (mm)
  • AC Input
    AC 220V ±10% / 50Hz
  • Input Active Power
    40w
  • Resolution
    AD:24bit;DA:24bit
  • Input Impedance
    ≥1GΩ
  • Voltage

    Constant Voltage Range Control:10mV~5V;

    Minimum Discharge Voltage:-5V;

    Accuracy:±0.01% of FS;

    Stability:±0.02% of FS;

  • Current

    Output Range per Channel: 0.2μA~0.1mA;0.1mA~1mA;1mA~10mA;10mA~100mA;

    Accuracy: ± 0.01% of FS;

    Constant Voltage Cut-off Current: 0.1μA;1μA;1μA;0.1mA

    Stability: ± 0.02% of FS;

  • Power

    Single Channel Maximum Output Power: 0.5W; 

    Stability: ± 0.1% of FS;

  • Time

    Current Response Time (10%FS~90%FS)

    ≤1.5ms;

     

    Step Time Range

    ① ≤(365*24)h/step;② Time Format Support  00:00:00:00(h, min, s, ms)

  • Data Record

    Data Recording Conditions

    Minimum Time Interval: 10ms; 

    Minimum Voltage Interval: 5mV; 

    Minimum Current Interval: 0.1μA;1μA;10μA;0.1mA


    Data Acquisition Frequency

    100Hz;

  • Charge

    Charging Mode: Constant Current Charging, Constant Voltage Charging, Constant Current/ Constant Voltage Charging, Constant Power Charging;

    Cut-off Conditions: Voltage, Current, Relative Time, Capacity, -△V;


  • Discharge

    Discharge Mode: Constant Current Discharge, Constant Voltage Discharge, Constant Current/ Voltage Discharge, Constant Power Discharge and Constant Resistance Discharge;

    Cut-off Conditions: Voltage, Current, Relative Time, Capacity;

  • Pulse Mode

    Charging: Constant Current Mode, Constant Power Mode;

    Discharge: Constant Current Mode, Constant Power Mode;

    Minimum Pulse Width: 500ms;

    Pulse Number: a single pulse step supports 32 different pulses;

    Continuous Switching between Charging and Discharging: One Pulse Step can Realize Continuous Switching from Charging to Discharging;

    Cut-off Conditions: Voltage, Relative time;

  • DCIR Test
    Support DCIR calculating with user defined point.
  • Cycle Life

    Cycle Test Range:1~65535次;

    Single Cycle Steps:254;

    Loop nesting: it has the function of nested loop and supports up to 3 levels of nesting;


  • Protect

    Power down Data Protection;

    Function of Offline Test;

    The safety protection conditions can be set, and the setting specs include: upper voltage limit, lower voltage limit, upper current limit, lower current limit, upper capacity limit and delay time.


  • IP Protection Degree
    Protection Degree: IP20
  • Channel Characteristics
    Constant current and constant voltage adopt double closed loop structure.
  • Channel Control Mode
    Independent Control
  • Voltage and Current Sampling
    Four Wire Connection
  • Noise
    <45dB
  • Database
    Adopting MySQL database to centralized manage testing data.
  • Communication Mode of Upper Computer
    Based on TCP/ IP Protocol
  • Server Operating System
    Windows 7 or Windows 10
  • Data Output Mode
    EXCEL2003,2010、TXT
  • Server Disk Configuration
    500GB
  • Communication Interface
    Network Interface
  • Leakage Current
    0.005μA
  • Total Number of Channels
    8
  • Working Temperature Range
    0℃~40℃(Within the range of 25±10℃, accuracy drift is 0.005% of FS /℃)
  • Storage Temperature Range
    -10℃~50℃
  • Relative Humidity Range of Working Environment
    ≤70% RH (There is no condensation of moisture.)
  • Relative Humidity Range of Storage Environment
    ≤80% RH (There is no condensation of moisture.)
  • Clamp Type
    Upper and Lower Clamps
  • Clamp Picture
    A705-P-QS-B
    PCB Coin Cell Clamps
    A705-P5-45B+(-)
Neware and Cookies
This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies.Read our private policy