Size of Each Unit (W*D*H)(mm)W*D*H: 443*615*125(mm)
-
AC InputAC 220V +10%/ -20%, 50Hz;
-
ResolutionAD: 16bit;DA: 16bit
-
Input Impedance≥ 500MΩ 开机状态(关机时漏电流为100uA)
-
Input Active Power500w
-
Channel Characteristics四量程 宽动态范围 高速采样 高精度
-
Channel Control ModeIndependent Control
-
Voltage
Constant Voltage Range Control
Charging:0.7 V~5V;
Discharging:0.7 V~5V (Clamp Line Length: 0.5m );
Minimum Discharge Voltage
0.7 V (Clamp Line Length: 0.5m );
Accuracy
± 0.02% of FS;
Stability
± 0.005% of FS;
-
Current
Output Range per Channel
Range 1: 0.1uA---150uA;
Range 2:150uA---5mA;
Range 3:5mA---150mA;
Range 4:150mA---5A;
Accuracy
± 0.02% of FS;
Constant Voltage Cut-off Current
2μA;0.01mA;0.02mA;
Stability
± 0.005% of FS;
-
Power
Single Channel Maximum Output Power: 25W;
Stability: ± 0.01% of FS;
-
Time
Current Response Time (10%FS~90%FS)
≤100μS (10% to 90% or 90% to 10%);
Step Time Range
≥10ms
-
Data Record
Data Recording Conditions
Time △t:≥10ms;
Voltage △U:≥1mV;
Current △I:≥100nA;
Data Acquisition Frequency
100Hz; (Continuous Charge & Discharge Mode);
-
Charge
Charging Mode: Constant Current Charging, Constant Voltage Charging, Constant Current/ Constant Voltage Charging, Constant Power Charging, Constant Resistance Charging;
Cut-off Conditions: Voltage, Current, Relative Time, Capacity, Energy, Power;
-
Discharge
Discharge Mode: Constant Current Discharge, Constant Power Discharge and Constant Resistance Discharge;
Cut-off Conditions: Voltage, Current, Relative Time, Capacity;
-
DCIR TestSupport DCIR Working Steps
-
Cycle Life
Cycle Test Range: 1~65535 times;
Single Cycle Steps: 254;
Loop nesting: it has the function of nested loop and supports up to 4 levels of nesting;
-
Protect
Software Protection
Power down Data Protection;
Function of Offline Test;
The safety protection conditions can be set, and the setting specs include: upper voltage limit, lower voltage limit, upper current limit, lower current limit, upper capacity limit, Voltage Fluctuation Protection, Current Fluctuation Protection;
-
Voltage and Current SamplingFour Wire Connection
-
DatabaseAdopting MySQL database to centralized manage testing data.
-
Communication Mode of Upper ComputerBased on TCP/ IP Protocol
-
Data Output ModeEXCEL2003,2010、TXT
-
Communication InterfaceNetwork Interface
-
Total Number of Channels8
-
Clamp TypeUniversal Clamps
-
Server Operating SystemWindows 7 64Bit
-
Working Temperature Range25±5℃ (Accuracy Assurance),25±20℃ (Extreme Working Temperature);
-
Storage Temperature Range0~60℃
-
Relative Humidity Range of Working Environment≤70% RH(There is no condensation of moisture.)
-
Relative Humidity Range of Storage Environment≤80% RH(There is no condensation of moisture.)
-
Clamp Picture