High Precision Battery Testing
CT-9008
  • Each Unit Size (W*D*H)(mm)
    W*D*H:443*615*125(mm)
  • AC Input
    AC 220V +10%/ -20%, 50Hz;
  • Input Active Power
    500W
  • Resolution
    AD:16bit;DA:16bit
  • Input Impedance
    ≥ 500MΩ 开机状态(关机时漏电流为100uA)
  • Voltage

    Voltage Range per Channel

    充电:0.7 V~5V;放电:0.7 V~5V(0.5m夹具线长);


    Minimum Discharge Voltage

    0.7V(0.5m夹具线长);


    Accuracy

    ± 0.1% of FS;


    Stability

    ± 0.005% of FS;

  • Current

    Output Range per Channel

    Range 1: 0.1μA~150μA;

    Range 2: 150μA~5mA;

    Range 3: 5mA~150mA;

    Range 4: 150μA~5A;


    Accuracy

    ± 0.02% of FS;


    Constant Voltage Cut-off Current

    Range 1: ±30nA;

    Range 2: ±1μA;

    Range 3: ±30μA;

    Range 4: ±1mA;


    Stability

    ± 0.005% of FS;

  • Power

    Single Channel Maximum Output Power: 25W;

    Stability: ± 0.01% of FS;

  • Time

    Current Response Time (10%FS~90%FS)

    ≤100μs


    Step Time Range

    ≥10ms

  • Data Acquisition

    Data Acquisition Conditions

    Time△t: ≥10ms;

    Voltage△U: ≥1mV;

    Current△I: ≥100nA;


    Data Acquisition Frequency

    1000Hz (Continuous Charge and Discharge Mode);

  • Charge

    Charging Mode: Constant Current Charging, Constant Voltage Charging, Constant Current/ Constant Voltage Charging, Constant Power Charging;

    Cut-off Conditions: Voltage, Current, Relative Time, Capacity, Energy, Power;


  • Discharge

    Discharge Mode: Constant Current Discharge, Constant Power Discharge and Constant Resistance Discharge;

    Cut-off Conditions: Voltage, Current, Relative Time, Capacity;

  • Cycle Life

    Cycle Test Range: 1~65535 times;

    Number of Single Cycle Steps: 255;

    Loop Nesting: It has the Function of Nested Loop and Supports Up to 4 Levels of Nesting;

  • DCIR Test
    Support DCIR Step.
  • Protect

    Power down Data Protection;

    The safety protection conditions can be set, and the setting specs include: Voltage Upper Limit, Voltage Lower Limit, Current Upper Limit, Current Lower Limit, Capacity Upper Limit Protection, Voltage Fluctuation Protection, Current Fluctuation Protection;

    Function of Offline Test.

  • Channel Characteristics
    4 Range, Wide Dynamic Range, High Speed Sampling, High Precision
  • Channel Control Mode
    Independent Control
  • Voltage and Current Sampling
    Four Wire Connection
  • Database
    Adopting MySQL database to centralized manage testing data.
  • Communication Mode of Upper Computer
    Based on TCP/ IP Protocol (100M Ethernet)
  • Server Disk Configuration
    500GB
  • Server Operating System
    Windows 7  64Bit
  • Leakage Current
    0.005mA
  • Data Output Mode

    EXCEL 2003, 2010; 

    TXT.

  • Communication Interface
    Network Interface
  • Total Number of Channels
    8
  • Working Temperature Range
    25±5℃(精度保证),25±20℃(极限使用温度);
  • Storage Temperature Range
    0~60℃
  • Relative Humidity Range of Working Environment
    ≤70% RH(There is no condensation of moisture.)
  • Relative Humidity Range of Storage Environment
    ≤80% RH(There is no condensation of moisture.)